In the fields of medical research and pathological analysis, negative pressure autopsy tables are crucial tools. To ensure they are always in the best working condition and provide reliable support for scientific research and medical work, correct maintenance and care are indispensable. Now, let's delve deeply into the maintenance and care strategies for negative pressure autopsy tables.
I. Key Points of Daily Maintenance
1. After each use, be sure to clean the tabletop promptly. Use a disinfected damp cloth to wipe off residues such as blood and tissue fluid to avoid bacterial growth.
2. Check the sealing of the tabletop to ensure there are no gaps or damages that might affect the negative pressure effect.
II. Regular Maintenance Items
1. Filter Replacement
- Filters are key to ensuring air quality. Generally, primary filters should be replaced every 1-2 months, medium-efficiency filters every 3-6 months, and high-efficiency filters every 6-12 months. However, the specific replacement frequency should be flexibly adjusted based on the usage environment and frequency.
- When replacing the filters, first turn off the equipment power and operate correctly as per the manual to prevent improper installation from affecting the filtration effect.
2. Fan Inspection
- Regularly check whether the fan operates normally, and if there are any abnormal noises or vibrations.
- Clean the fan blades and vents of the fan to prevent dust accumulation from affecting the fan performance.
3. Pipeline Cleaning
- Check the interior of the pipelines connected to the negative pressure autopsy table for dirt and blockages at regular intervals.
- Professional pipeline cleaning tools can be used for cleaning to ensure the pipelines are unobstructed.
4. Electrical System Maintenance
- Inspect whether the wires are damaged and if the plugs are loose.
- Conduct functional tests on the buttons and indicator lights on the control panel to ensure they work properly.
III. Handling Special Situations
1. If the negative pressure effect weakens, first check if the filters need to be replaced and if there are any leaks in the pipelines.
2. In case of equipment failure or abnormal situations, stop using it immediately and contact professional maintenance personnel for repair.
IV. Maintenance Tips
1. Lay disposable protective pads on the tabletop, which not only facilitates cleaning but also protects the tabletop.
2. Establish a maintenance record file, detailing the time, content of each maintenance, and replaced components for easy tracking and management.
For example, there was once a medical institution that neglected the regular maintenance of the negative pressure autopsy table, resulting in a clogged filter and a weakened negative pressure effect, affecting work efficiency and the accuracy of experimental results.
Correct maintenance and care are the keys to ensuring the normal operation and extending the service life of negative pressure autopsy tables. Hope everyone can attach importance to it and let the negative pressure autopsy tables better serve our medical cause.